IDEAL INDUSTRIES Thermometer 61 633 User Manual

IDEAL INDUSTRIES INC.  
TECHNICAL MANUAL  
MODELS: 61-633  
61-635  
The Service Information provides the following information:  
• Precautions and safety information  
• Specifications  
• Basic maintenance (cleaning, replacing the battery and fuses)  
• Performance test procedures  
• Calibration and calibration adjustment procedures  
Form Number: TM61633-5  
Revision: 5. Date: December 2006  
Form Number TM61633-5  
Rev 5 Dec 2006  
 
Page 1  
Introduction  
Warning  
To avoid shock or injury, do not perform the verification tests or calibration  
procedures described in this manual unless you are qualified to do so.  
The information provided in this document is for the use of qualified personnel only.  
Caution  
The 61-630 serials contain parts that can be damaged by static discharge.  
Follow the standard practices for handling static sensitive devices.  
For additional information about IDEAL INDUSTRIES and its products,  
and services, visit IDEAL INDUSTRIES web site at:  
Precautions and Safety Information  
Use the Meter only as described in the Users Manual. If you do not do so, the  
protection provided by the Meter may be impaired. Read the “Safety Information”  
page before servicing this product. In this manual, a Warning identifies conditions  
and actions that pose hazard (s) to the user; a Caution identifies conditions and  
actions that may damage the Meter or the test instruments.  
The Symbols  
The symbols used on the Meter and in this manual are explained in Table A.  
Table A. The Symbols  
Symbol Meaning  
Alternating signal  
Symbol Meaning  
Battery  
Direct signal  
Earth ground  
Refer to the manual.  
Important information.  
Fuse  
Take appropriate precautions.  
Hazardous voltage may be  
present  
Double insulated  
CAT III IEC over voltage Category III  
Form Number TM61633-5  
Rev 5 Dec 2006  
 
Page 2  
SAFETY  
Review the following safety precautions to avoid injury and prevent damage to this product or any products  
connected to it. To avoid potential hazards, use the product only as specified.  
CAUTION.  
These statements identify conditions or practices that could result in damage to the equipment or other property.  
WARNING.  
These statements identify conditions or practices that could result in personal injury or loss of life.  
Specific precautions  
Use proper Fuse. To avoid fire hazard, use only the fuse type and rating specified for this product.  
Do not operate without covers. To avoid personal injury, do not apply any voltage or current to the product  
without the covers in place.  
Electric overload. Never apply a voltage to a connector on the product that is outside the range specified for that  
connector.  
Avoid electric shock. To avoid injury or loss of life, do not connect or disconnect probes or test leads while they  
are connected to a voltage source.  
Do not operate in wet/damp conditions. To avoid electric shock, do not operate this product in wet or damp  
conditions.  
Form Number TM61633-5  
Rev 5 Dec 2006  
 
Page 3  
SPECIFICATIONS  
All specifications are warranted unless noted typical and apply to models 61-633 and 61-635.  
Stated accuracies are at 23° ±5°C at less than 80% relative humidity and without the battery indicator displayed.  
General specifications  
Characteristics  
LCD display digits  
Bargraph segments  
Display count  
Description  
4¾ or 3¾  
80 Segment Graph  
40,000 or 4,000  
Numeric update rate  
2 times / sec (40.000 count)  
4 times / sec (4,000 count)  
20 times/sec  
Bargraph  
Polarity display  
Automatic  
Overrange display  
Low voltage indicator  
Automatic power-off time  
Power source  
OL is displayed  
is indicated  
Automatic backlit off = 15 minutes  
One 9V dry cell battery  
Maximum input voltage  
Maximum floating voltage  
1000V (750V AC) CAT II, Between V and COM  
1000V (750V AC) CAT II, Between any terminal and earth  
ground  
Maximum input current  
400mA between mA and COM  
10A continuous between A and COM  
(20A for 30 seconds, between A and COM)  
Maximum open circuit Voltage (current inputs) 600V between A and COM and between mA and COM  
Overload protection mA connector  
A connector  
1 A (600V) fast blow fuse  
15A (600V) fast blow fuse  
V connector  
1100 Vp, V , V AC+DC  
850 Vp, mV , mV , AC+DC  
LVΩ, Ω, Hz,  
0.1 x (Spec. Accuracy) 1°C, < 18 C or > 28°C.  
,
,
, %DF, °C, °F  
Temperature Coefficient  
Battery Life  
100 hours typical (alkaline)  
Form Number TM61633-5  
Rev 5 Dec 2006  
 
Page 4  
Measurement Characteristics  
All at 23°C ±5°C, < 80% Rh. Specifications are expressed as ±([% of reading] + [number of digits]).  
Multiply accuracy Digits by l0 in 40,000 count Mode.  
DC VOLTAGE:  
DCV  
40mV  
400mV  
61-633  
61-635  
± (0.20%+8d)  
± (0.20% + 2d)  
± (0.20% + 2d)  
± (0.06% + 8d)  
± (0.06% + 2d)  
± (0.06% + 1d)  
4V, 40V, 400V, 1000V  
AC VOLTAGE:  
ACV  
61-633  
61-635  
400mV  
40Hz~ 100Hz  
100Hz ~ 1KHz  
4V  
± (1.20% + 5d)  
± (2.00% + 5d)  
± (0.70% + 5d)  
± (1.00% + 5d)  
40Hz ~ 100Hz  
100Hz ~ 1KHz  
1KHz ~ 10KHz  
10KHz ~ 20KHz  
20KHz ~ 50KHz  
50KHz ~ 100KHz  
40V  
± (1.00% + 5d)  
± (2.00% + 5d)  
± (3.00% + 6d)  
N.A  
± (0.70% + 5d)  
± (1.00% + 5d)  
± (2.00% + 6d)  
± (3.00% + 7d)  
± (5.00% + 8d)  
± (10.00% + 10d)  
N.A  
N.A  
40Hz ~ 100Hz  
100Hz ~ 1KHz  
1KHz ~ 10KHz  
10KHz ~ 20KHz  
20KHz ~ 50KHz  
50KHz ~ 100KHz  
400V  
± (1.20% + 5d)  
± (2.00% + 5d)  
± (3.00% + 6d)  
N.A  
± (0.70% + 5d)  
± (1.00% + 5d)  
± (2.00% + 6d)  
± (3.00% + 7d)  
± (5.00% + 8d)  
± (10.00% + l0d)  
N.A  
N.A  
40Hz ~ 100Hz  
100Hz ~ 1KHz  
1KHz ~ 10KHz  
10KHz ~ 20KHz  
20KHz ~ 50KHz  
750V  
± (1.20% + 5d)  
± (2.00% + 5d)  
±(3.00% + 6d)  
N.A  
± (0.70% + 5d)  
± (1.00% + 5d)  
± (2.00% + 6d)  
± (3.00% + 7d)  
± (5.00% + 8d)  
N.A  
40Hz ~ 100Hz  
100Hz ~ 1KHz  
±(1.20% + 5d)  
± (2.00% + 5d)  
± (0.70% + 5d)  
± (1,00% + 6d)  
Bandwidth  
40Hz ~ 10KHz  
40Hz ~ 100 KHz  
dBm (typical): -l5dBm to +55 dBm (0 dBm = lmW into 600Ω).  
dBv (typical): -80 dBv to + 50 dBv (0 dBv = 1 Vrms).  
Note: (ACV only)  
Add additional 40Digits for reading under 30% of range.  
Specifications exclude under 20% of range for 20KHz ~ 100KHz.  
Resolution: 1µV in the 400mV range  
Input Impedance: 10MΩ, <l00pF  
Overload Protection: 1000V dc, 750V rms  
AC Conversion Type: AC Coupled True RMS responding  
AC + DC Volts: Same as AC(RMS) +1.00% + 8d  
Crest Factor: +1.5% addition error for C.F. 1.4 to 3  
+3.0% addition error for C.F. from 3 to 4  
Form Number TM61633-5  
Rev 5 Dec 2006  
 
Page 5  
DC CURRENT:  
DCA  
40mA, 400mA, 4A, 10A  
61-633  
± (0.50% + 4d)  
61-635  
±(0.20% + 4d)  
AC CURRENT:  
ACA  
61-633  
61-635  
40mA, 400mA, 4A, 10A  
Bandwidth  
±(1.20% + 8d)  
40Hz ~ 400Hz  
± (0.80% + 8d)  
40Hz ~ 400Hz  
Range:  
Resolution:  
40mA, 400mA, 4A, 10A  
1µA in the 40mA range  
Burden Voltage: 800mV max. for mA input, 1V max. for A input  
AC Conversion Type: AC Coupled True RMS responding  
Input Protection: Equipped with High Energy Fuse  
1A, 600V, IR 10KV fuse (Bussmann BBS-1 or equivalent) for mA input  
15A, 600V, IR 100KV fuse (Bussmann KTK 15 or equivalent) for A input  
AC + DC Current: Same as AC(RMS) + 1.00% + 8d  
C.F.: Same as ACV  
PEAK HOLD: +[± (0.7% + 20)] additional error for >10% of full scale  
RESISTANCE:  
OHM  
4000, 4K 0  
40KQ, 400K0  
4M0  
61-633  
61-635  
± (0.50% + 2d)  
± (0.50% + 2d)  
± (0.50% + 4d)  
± (5.00% + 5d)  
± (0.30% + 2d)  
± (0.30% + 2d)  
± (0.30% + 4d)  
± (5.00% + 5d)  
40M0  
LV OHM  
4K0 ,40K0 400K 0  
4M0  
61-633  
61-635  
± (1.00% + 2d)  
± (1.00% + 4d)  
± (7.00% + 5d)  
± (0.60% ÷ 2d)  
± (0.60% + 4d)  
± (7.00% + 5d)  
40M0  
Resolution: 0.01 Ω in the 4000 range  
Open Circuit Voltage: 3.3V  
Open Circuit Low Voltage: 0.6V  
Input Protection: 600V rms  
CONTINUITY CHECK  
Continuity Threshold: Approximately 50Ω  
Continuity Indicator: 2KHz Tone Buzzer  
Input Protection: 600V rms  
DIODE TEST  
Test Current: 1.1 mA (Typical)  
Open Circuit Voltage: 3.3V DC (max)  
Input Protection: 600V rms  
Form Number TM61633-5  
Rev 5 Dec 2006  
 
Page 6  
CAPACITANCE  
Capacitance  
61-633  
61-635  
4nF, 40nF  
400nF, 4µF  
40µF, 400µF  
4mF, 10mF  
± (1 .90% + 20d)  
± (0.90% + 20d)  
± (2.90% + 20d)  
± (3.90% + 20d)  
± (1 .90% + 20d)  
± (2.90% + 20d)  
Note: For best measurements, use REL mode in the nF ranges.  
Range: 4nF, 40nF, 400nF, 4µF, 40µF, 400µF, 4mF, l0mF  
Resolution: lpF in the 4nF range  
Input Protection: 600V rms  
FREQUENCY COUNTER  
Range: 400Hz, 4KHz, 40KHz, 400KHz, 4MHz  
Resolution: 0.01Hz in the 400Hz range  
Accuracy: ± (0.01% + 1d)  
Sensitivity: 0.5Vp-p, for 15Hz ~ 1 MHz, 1 Vp-p, for 1MHz ~ 4MHz  
Min. Frequency: 15Hz.  
Input Protection: 600V rms  
DUTY FACTOR  
Range: 20% ~ 80%  
Resolution: 0.1%  
Accuracy: ±6d (15Hz ~ 10KHz, 5Vp-p)  
Temperature  
Temperature  
-50°C ~ 1200°C  
-100°C ~ -50°C  
-200°C ~ 100°C  
61-633  
N.A  
N.A  
61-635  
1°C + ld  
2°C + 1d  
3°C + 1d  
N.A  
Physical Characteristics  
Characteristics  
Description  
200mm x 90mm x 42mm  
212mm x 100mm x 55mm(with holster)  
Dimensions (H x W x D)  
Weight (with battery)  
With holster  
0.4Kg  
0.6Kg  
Environmental Characteristics  
Temperature operating  
Non-Operating  
Description  
0 to + 50  
-20 to + 60  
Humidity (operating)  
Altitude Operating  
Non-Operating  
< 80% R.H.  
2,222 m (7290 ft.)  
12,300 m (40354 ft.)  
MIL-T-28800E TYPE II Class 5  
2.66gRMS, 5 to 500 Hz, 3axes  
(10 minutes each)  
Vibration & shock Operating  
Dust / Water Protection IP Rating  
Indoor Use  
IP 64  
Form Number TM61633-5  
Rev 5 Dec 2006  
 
Page 7  
Certifications and compliances  
Safety  
Designed to IEC 1010-1, UL3111-1 and CSA specifications  
1000V DC Category II  
600V DC Category III  
Input rating  
750V AC Category II  
600V AC Category III  
CAT III: Distribution level mains, fixed installation.  
CAT II: Local level mains, appliances, portable equipment.  
Over voltage category  
Pollution Degree 2  
CAT I: Signal level, special equipment or parts  
of equipment, telecommunication, electronics.  
Do not operate in environments where conductive Pollutants may be  
present.  
Meets the intent of Directive 89/336/EEC for  
Electromagnetic Compatibility and Low Voltage  
Directive 73/23/EEC for Product Safety. Compliance was  
demonstrated to the following specifications as listed in the  
official Journal of the European Communities:  
En 55011 Class A: Radiated and Conducted Emissions.  
EN 50082-1 Immunity:  
EC Declaration of  
Conformity  
IEC 801-2 Electrostatic Discharge  
IEC 801-3 RF Radiated  
EN 61010-1 Safety requirements for electrical equipment  
for measurement, control, and laboratory use.  
Form Number TM61633-5  
Rev 5 Dec 2006  
 
Page 8  
Required Equipment  
Required equipment is listed in Table B. If the recommended  
models are not available, equipment with equivalent specifications may be used.  
Repairs or servicing should be performed only by qualified personnel.  
Table B. Required Equipment  
Recommended  
Equipment Required Characteristics  
Model  
Calibrator  
AC Voltage Range: 0-750V ac  
Accuracy: ±0.07% (Basic)  
Frequency Range: 40 ~ 100KHz  
Accuracy: ± 2%  
Fluke 5500 or  
Wavetek 9100  
Calibrator or  
equivalent  
DC Voltage Range: 0-1000V dc  
Accuracy: ±0.006% (Basic)  
Current Range: 0 ~ 10A  
Accuracy:  
AC (40Hz to 1KHz): ±0.08% (Basic)  
DC: ± 0.02% (Basic)  
Frequency Source:  
5.00Hz ~ 1.0000MHz  
Accuracy: ±0.001%  
Amplitude:  
0.5V p-p ~ 1.0V p-p (square wave)  
Accuracy: ± 5%  
range: 1~ 2G  
Accuracy: ±0.03% (Basic)  
Capacitance Range: 1PF ~ 40mF  
Accuracy: ±0.10% (Basic)  
Temperature Range: -200 ~ 1200  
Accuracy: ±0.3 (Basic)  
Form Number TM61633-5  
Rev 5 Dec 2006  
 
Page 9  
Basic Maintenance  
Warning  
To avoid shock, remove the test leads and any input signals before opening  
the case or replacing the battery or fuses.  
Opening the Meter Case  
Caution  
To avoid unintentional short circuit, always place the uncovered Meter  
assembly on a protective surface. When the case of the Meter is open, circuit  
connections are exposed.  
To open the Meter case, refer to Figure 1 and do the following:  
1. Disconnect test leads from any live source, turn the rotary switch to OFF, and remove the  
test leads from the front terminals.  
2. Remove the battery door by using a Phillips-head screwdriver to remove the screw to the  
battery door.  
3. The case bottom is secured to the case top by three screws and two internal snaps  
(at the LCD end). Use a Phillips-head screwdriver to remove the three screws.  
4. Un-snap the Battery and table the battery snap out of the battery room.  
Note  
The gasket between the two case halves is sealed to, and must remain with, the case bottom. The case top lifts  
away from the gasket easily. Do not damage the gasket or attempt to separate the case bottom from the gasket.  
5. Hold the Meter display side up.  
6. Lifting up on the input terminal end, disengage the case top from the gasket.  
7. Gently unsnap the case top at the display end.  
Battery Replacement (refer to Figure 1)  
1. Disconnect the test leads from any circuit under test and turn off meter.  
2. Remove the test leads from meter.  
3. Loosen the screw from the battery cover on bottom case.  
4. Remove battery cover.  
5. Install a new battery after removing the original one.  
6. Assemble battery cover onto bottom case with screwdriver and the screw  
described in step 3.  
Testing Fuses (FS1 and FS2)  
To test the internal fuses of the Meter, refer to Figure 1 and do the following:  
1. Turn the rotary selector switch to the mA position for 1A fuse test FS1 or A  
position for 15A fuse test FS2.  
2. To test FS1, plug a test lead into mA input terminal, and turn the rotary selector to the  
mA position.  
2.1 Fuse is OK if display shows normal functional graphics.  
2.2 Fuse is defective if “FUSE” is displayed, and the built-in beeper alarms.  
3. To test FS2, plug a test lead into A input terminal, and turn the rotary selector to the  
A position.  
3.1 Fuse is OK if the display shows normal functional graphics  
3.2 Fuse is defective if “FUSE” is displayed, and the built-in beeper alarms.  
Form Number TM61633-5  
Rev 5 Dec 2006